No products in the cart.
- Simultaneous Power & Power quality measurements
Power/ Harmonics/ Waveform/ Power quality are recorded at all CHs. (Voltage: 3ch, Current 4ch) - Helpful support functions
Quick Start Guide, Wiring check and Sensor detection for easy and reliable measurement - Measurement with high accuracy
Guaranteed accuracy: ±0.3%rdg (energy),
±0.2%rdg (voltage/ current)
Complies with the International Standard IEC 61000-4-30 Class S
and the European Standard EN50160 - Remote monitoring on PC and Android device
Remote checking of measurement in real-time is possible via Bluetooth communication.
Recorded data can be saved in the supplied SD card.
EN50160 report can be generated after survey by PC software. - Various Clamp Current Sensors
Various types of clamp and flexible sensors are available:
from 1000mA Range up to 3000A Range and Earth leakage measurements - Energy consumption check on site
Trend and demand graphs for easy recognition.
TFT color display with high resolution. - IEC 61010-1 CAT IV 300V, CAT III 600V, CAT II 1000V
Specification | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Wiring connections | 1P2W, 1P3W, 3P3W, 3P4W | |||||||||
Measurements and parameters | Voltage, Current, Frequency, Active power, Reactive power, Apparent power, Active energy, Reactive energy, Apparent energy, Power factor (cos θ), Neutral current, Demand, Harmonics, Quality (Swell/Dip/Interruption, Transients/Over voltage, Inrush current, Unbalance rate), Capacitance calculation for PF correction unit, Flicker | |||||||||
Voltage (RMS) | ||||||||||
Range | 600.0 / 1000V | |||||||||
Accuracy | ±0.2%rdg±0.2%f.s. (sine wave, 40 - 70Hz) | |||||||||
Allowable input | 1 - 120% of each range (rms). 200% of each range (peak) | |||||||||
Display range | 0.15 - 130% of each range | |||||||||
Crest factor | 3 or less | |||||||||
Sampling speed of Voltage transient | 24µs | |||||||||
Current (RMS) | ||||||||||
Range | ||||||||||
8128 | ( | 50A | type ) | : | 5 / 50A / AUTO | |||||
8127 | ( | 100A | type ) | : | 10 / 100A / AUTO | |||||
8126 | ( | 200A | type ) | : | 20 / 200A / AUTO | |||||
8125 | ( | 500A | type ) | : | 50 / 500A / AUTO | |||||
8124/8130 | ( | 1000A | type ) | : | 100 / 1000A / AUTO | |||||
8146/8147/8148 | ( | 10A | type ) | : | 1 / 10A / AUTO | |||||
8129 | ( | 3000A | type ) | : | 300 / 1000 / 3000A | |||||
Accuracy | ±0.2%rdg±0.2%f.s. + accuracy of clamp sensor (sine wave, 40 - 70Hz) | |||||||||
Allowable input | 1 - 110% of each range (rms). 200% of each range (peak) | |||||||||
Display range | 0.15 - 130% of each range | |||||||||
Crest factor | 3 or less | |||||||||
Active power | ||||||||||
Accuracy | ±0.3%rdg±0.2%f.s. + accuracy of clamp sensor (power factor 1, sine wave, 40 - 70Hz) | |||||||||
Influence of power factor | ±1.0%rdg (reading at power factor 0.5 against power factor 1) | |||||||||
Frequency meter range | 40 - 70Hz | |||||||||
Power supply (AC Line) | AC100 - 240V / 50 - 60Hz / 7VA max | |||||||||
Power supply (DC battery) | Alcaline size AA battery LR6 or Ni-MH (HR-15-51) × 6 Battery life approx. 3h (LR6, Backlight OFF) | |||||||||
Internal memory | FLASH memory (4MB) | |||||||||
PC card interface | SD card (2GB) | |||||||||
PC communication interface | USB Ver2.0, Bluetooth Ver2.1 + EDR Class2 | |||||||||
Display | 320 × 240 (RGB) Pixel, 3.5inch color TFT display | |||||||||
Display update period | 1 sec | |||||||||
Temperature and humidity range | 23±5ºC, less than 85% RH (without condensation) | |||||||||
Operating temperature and humidity range | 0 - +45ºC, less than 85% RH (without condensation) | |||||||||
Storage temperature and humidity range | -20 - +60ºC, less than 85% RH (without condensation) | |||||||||
Applicable standards | IEC 61010-1 CAT IV 300V, CAT III 600V, CAT II 1000V Pollution degree 2 IEC 61010-2-030, IEC 61010-031, IEC 61326, EN 50160 IEC 61000-4-30 Class S, IEC 61000-4-15, IEC 61000-4-7 | |||||||||
Dimension | 175(L) × 120(W) × 68(D) mm | |||||||||
Weight | Approx. 900g | |||||||||
Included accessories | ||||||||||
7141B | (Voltage test lead) | |||||||||
7170 | (Power cord) | |||||||||
7219 | (USB cable) | |||||||||
8326-02 | (SD card 2GB) | |||||||||
9125 | (Carrying case for KEW 6315) | |||||||||
9135 | (Carrying case for KEW 6315-03) | |||||||||
Input terminal plate × 6, KEW Windows for KEW 6315 (software), Quick manual, Alkaline size AA battery (LR6) × 6 | ||||||||||
Optional accessories | ||||||||||
8124, 8125, 8126, 8127, 8128 | (Load current clamp sensor) | |||||||||
8129, 8130 | (Flexible clamp sensor) | |||||||||
8146, 8147, 8148 | (Leakage and Load current clamp sensor) | |||||||||
8312 | (Power supply adaptor) | |||||||||
9132 | (Magnetci carrying case) |