• Sorry, this product cannot be purchased.

Kyoritsu 1062

Note : Harga dapat berubah sewaktu-waktuAuthorized Distributor

High Accuracy, High Performance and Reliable Measurements

  • 0.02% basic DC accuracy
  • Large display with 50,000 counts
  • Dual display for double indication for AC and DC, V and Hz, etc.
  • True-RMS Measurements AC and AC+DC
  • Wide AC Frequency bandwidth from 10Hz to 100kHz *
  • True-RMS or MEAN value detection mode can be selected *
  • Fast Peak Hold response time of 250µs
  • Low-pass filter for motor drive measurements *
  • Low Power Ω for resistance measurements on sensible electronic circuits by a low and safe test current *

Safety design for industrial use

  • Complies with IEC 61010-1 CAT III 1000V, CAT IV 600V
  • Terminal shutter to prevent incorrect test leads’ insertion in current terminals
  • Very wide operating temperature range from -20 to 55ºC

Reliable support for data management

  • Large data internal memory
    KEW 1062: 10,000 data (logging), 100 data (manual)
    KEW 1061: 1,000 data (logging), 100 data (manual)
  • Download data and Live Monitoring on a PC via the USB interface
    (Option for USB Communication set)
KEW 1061 KEW 1062
Detection mode RMS MEAN/RMS (switch)
DC V 50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V (Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V]) |±0.02%rdg±2dgt (Basic accuracy)
AC V [RMS] 50.000*/500.00mV/5.0000/50.000/500.00/1000.0V(* Only for 1062) (Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V])
|±0.7%rdg±30dgt (Basic accuracy) |±0.4%rdg±30dgt (Basic accuracy)
AC V [MEAN] * -- 50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V (Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V]) |±1%rdg±30dgt (Basic accuracy)
DCV+ACV 5.0000/50.000/500.00/1000.0V (Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V])
|±1%rdg±10dgt (Basic accuracy) |±0.5%rdg±10dgt (Basic accuracy)
DC A 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A |±0.2%rdg±5dgt (Basic accuracy)
AC A [RMS] 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1%rdg±20dgt (Basic accuracy) |±0.75%rdg±20dgt |(Basic accuracy)
AC A [MEAN] * -- 500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A |±1.5%rdg±20dgt (Basic accuracy)
DCA+ACA 500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1.5%rdg±10dgt (Basic accuracy) |±1%rdg±10dgt (Basic accuracy)
Ω 500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ
|±0.1%rdg±2dgt (Basic accuracy) |±0.05%rdg±2dgt (Basic accuracy)
LowPower-Ω * -- 5.000/50.00/500.0kΩ/5.000MΩ |±0.2%rdg±3dgt (Basic accuracy)
Continuity buzzer 500.0Ω (The buzzer turns on for resistances lower than 100±50Ω)
Diode test 2.4000V |±1%rdg±2dgt Open curcuit voltage: <5V (Approx. 0.5mA Measuring Current)
Diode test 5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF
±1%rdg±5dgt (Basic accuracy)
Frequency 5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF
±1%rdg±5dgt (Basic accuracy)
DUTY 10 - 90% |±1%rdg
Temperature -200 - 1372ºC |±1%rdg±1.5ºC
-328 - 2501.6ºF |±1%rdg±2.0ºF
(with the use of K-type Temperature probe)
Measurement functions DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω *s
Other functions Data Hold (D/H), Auto Hold (A/H), Peak Hold * (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight
Withstand voltage 6,880V/5 sec.
Applicable standards IEC 61010-1 CAT IV 600V, CAT III 1000V Pollution degree 2, IEC 61010-031, IEC 61326-1 (EMC)
Dimensions 192(L)×90(W)×49(D) mm
Weight Approx. 560g (including batteries)
Accessories 7220A (Test leads), R6×4, Instruction manual,
8926 (Fuse [440mA/1000V])×1 (included)
(with the use of K-type Temperature probe)